One-point Calibration

A one-point slope calibration requires one known standard. For conductivity, use one
of the solutions in Appendix C. For dissolved oxygen, use 100% humidified air (see
“DO

2

Calibration” on page 7 for details).

1.

From the Sensor field, select Water Quality.

2.

From the Measurement field, select Dissolved Oxygen
(mg/L), Dissolved Oxygen (%), Conductivity (1X Probe),
or
Conductivity (10X Probe).

3.

Ensure that “Calibrate all similar measurements simulta-
neously” is not selected.

4.

From the Calibration Type menu, select 1 Point (Adjust Slope
Only)
, if it is not already selected.

5.

Place the probe in the known standard.

6.

Enter the known conductivity or DO

2

concentration in the

Standard Value field for Point 2.

7.

Observe the Present Sensor Measurement and wait until it
has stabilized. (For conductivity, ensure that the present mea-
surement is within the selected sensor range. If it is not, press a
button on the multi-sensor to select a higher range.)

8.

Click the Read From Sensor button for Point 2.

9.

Click OK to accept the new calibration.

Appendix B: EZ Screen

If you are using the Water Quality MultiMeasure Sensor with a PASPORT interface
connected to a computer, you can use EZ-Screen software instead of DataStudio.
When the sensor is connected to the computer through the interface (if DataStudio is
not running) a window automatically launches giving you the option of launching
EZ-Screen. In EZ-Screen, click the green arrow button to start data collection. Click
the Water Quality icon in the lower-right corner to cycle through temperature, dis-
solved oxygen, probe voltage, conductivity, and pH measurements.