Specifications – PASCO PS-2169 Water Quality MultiMeasure Sensor User Manual

Page 9

Advertising
background image

®

M o d e l N o . P S - 2 1 6 9

S p e c i f i c a t i o n s

9

Specifications

General

Component Sensors

Temperature, Conductivity, pH/ISE/ORP probe voltage, Dissolved Oxygen

Max. Sampling Rate

20 samples per second (for each component sensor)

Default Sampling
Rate

2 samples per second

Temperature

Range

−35 °C to +135 °C

Accuracy

±0.5 °C

Resolution

0.01 °C or better

Sensing Element

10 k

Ω thermistor located in probe tip

Conductivity

Range

three user-selectable ranges:
0 to 1000

µS/cm, 0 to 10000 µS/cm, 0 to 100000 µS/cm with included 10x probe

0 to 100

µS/cm, 0 to 1000 µS/cm, 0 to 10000 µS/cm with optional 1x probe

Accuracy

±10% of full scale without calibration

Resolution

better than 0.05% of full scale

pH/ISE/ORP probe voltage

Electrode Connector

standard BNC

Input Resistance

1012

Voltage Range

−2000 mV to +2000 mV

Voltage Resolution

0.1 mV

pH Range

0 to 14

pH Resolution

0.001

Dissolved Oxygen

Range

0 to 20 mg/L

Accuracy

±0.2 mg/L after calibration
±2 mg/L without calibration

Resolution

0.01 mg/L

Temperature Range

0 °C to 50 °C

Advertising