5.2 Dryers

Under normal conditions Perma Pure dryers require little maintenance and can last for several
years. However, if there is no pre-filter and tubing becomes clogged or saturated with water, dryer
may require cleaning or repair. When disassembling the dryer, note that end fittings on the PD-
SERIES dryer can be easily rotated on shell tube. This rotation should be avoided to prevent
twisting membrane tubes inside the shell.

Dryer Element Replacement:

1.

Take dryer assembly out of unit by unscrewing 4 screws on heater block, 4 screws on
bottom of unit along with 2 allen hex head screws.

2.

Loosen all compression fittings connecting to dryer.

3.

Refer to Appendix B for Dryer Element Replacement.

When connecting any fittings to purge ports of dryer, be sure not to tighten threaded fittings
more than 5 turns. Additional turns may cause fitting to damage element header just below.

5.3 Fuse

Fuse Replacement - The fuse is located on or in the control enclosure.

To remove the fuse:

1.

Turn cap marked “FUSE” counter-clockwise ¼ turn and the fuse holder will come out.

2.

Remove blown fuse and replace with one of equal amperage rating. The standard fuse is a
BUSS type AGC or equivalent.

3.

After installing correct replacement fuse, re-install fuse holder by pressing inward and
twisting 1/4 turn clockwise.