Ambi-GASS User Manual

4. Start-Up Procedure

4.1 Setup Check

1.

Check electrical, sample, purge and drain connections have been made.

2.

Turn on instrument and/or compressed air to system.

4.2 Purge Air Flow Adjustment

Purge air to dryer is controlled by regulator and flowmeter combination on control panel.
Pressure setting are as follows:

1.

With optional filter drain – Set pressure to aprox. 10 psig.

2.

Without filter drain – Set pressure to aprox. 20 psig

3.

Adjust purge air flow to 1 1/2 to 2 times sample flow.

4.3 Purge Air Eductor (optional)

Adjust eductor regulator until vacuum gauge reads about 2-5 inches of mercury.
Re-adjusted each time changes are made to purge flow rate or purge pressure.