Roubleshooting – Perma Pure Baldwin 3300 FLow Control Drawer User Manual

Page 13

Advertising
background image

G:

T

ROUBLESHOOTING

Symptom Check

Action

No LED display

Not plugged in
Power switch is off
Fuse blown
Power supply defect
12 VDC red-black
Check jumper JP1


Ribbon cable loose
Defective control board

Plug in instrument
Turn on power
Replace fuse

Replace power supply
Jumper 1-2
Jumper 3-4
Jumper 5-6
Tighten ribbon cables
Replace control board

Manual Control (Local)
Switches have no effect
Inadequate sample flow
No system cal gas flow








No direct cal gas flow

Remote/Local switch in remote
Low sample pressure

No cal gas from bottle
Cal gas solenoid not operating

Flow drawer regulator set <10psig
Cal gas Flowmeter closed
Block/bleed solenoid constantly
vents
System cal gas solenoid valve
defective

Direct cal gas solenoid valve
defective

Select local
Adjust sample pressure
Cylinder to 15 psig
Check cabling
Replace cal gas solenoid
Replace control board
Set regulator to 10 psig
Adjust Flowmeter
Replace solenoid SV9

Replace solenoid SV10


Replace solenoid SV11
Check cables and connector
boards for good connections

Computer Control (Remote)
Computer control has no effect








No computer sense

Remote/Local switch in local
Computer control wiring to
terminal block is wrong




Ribbon cable is loose
Defective control board
Computer sense wiring to terminal
block is wrong
Ribbon cable is loose

Select remote
See the computer control wiring
section in manual, also if using
a PLC for control make sure to
match your DC GND with the
Unit DC GND reference points
on pins 14 or 16.
Check cables.
Replace control board
See computer control wiring
section in manual
Check cables

For further service assistance, contact: