Zero−air generator, Start-up procedure – Perma Pure ZERO-AIR GENERATOR User Manual

Page 5

Advertising
background image


4

Form #240

Zero-Air Generator Manual

Rev 11/98

0

5

10

15

20

25

30

35

0

2

4

6

8

10

12

14

16

18

20

Flow Rate (L/min)

P

ressu

re (p

sig

)

Start-Up Procedure

1. Place the ZAG on a level,

dry surface.


2. Check to make sure the

power switch is in the ‘off’ position.


3. Plug the unit into a

standard 110V electrical outlet.


4. Open the flow control knob

fully (turn the knob counter-clockwise).

5. Connect the calibration gas line to the 1/4”

tube compression fitting located on the top
panel of the generator.


6. Turn the power switch to the ‘on’ position.

7. Adjust the flow control valve to the desired air
output. Maximum capacity is 18 lpm.

8. High purity air will be produced within 2
minutes after start-up.

Note:
There is a
trade off between
air pressure and
flow. High air
volumes will result
in lower air
pressure, and vice
versa. If low
pressure, low flow
applications are
encountered, a
pressure regulator
should be
connected between
the instrument and
the Zero-Air
Generator.


Figure 4 – Zero-Air Generator Performance Curve

3

REPLACE ABSORBENT WHEN COLOR
CHANGE REACHES THIS POINT

OUTLET FLOW

(LPM)

SYSTEM

PRESSURE

ZERO−AIR

OUTLET

5

10

30
25

15

20

25

10

0

5

15
psi 20

30

Figure 3A – ZAG Top Panel

ZERO−AIR

GENERATOR

CAUTION : READ INSTRUCTION

MANUAL BEFORE OPERATING

FUSE 3A FAST ACTING 5x20mm

SYSTEM POWER 110VAC 50/60HZ

8 EXECUTIVE DRIVE P.O. BOX 2105
TOMS RIVER, NEW JERSEY 08754

PERMA PURE INC.

Figure 3B – ZAG Bottom Panel