Perma Pure ZERO-AIR GENERATOR User Manual

Page 7

Advertising
background image


6

Form #240

Zero-Air Generator Manual

Rev 11/98

4.2 Final Filter Replacement

The final filter should be replaced each time the activated carbon/alumina is changed. The
filter is located immediately after the outlet of the CO catalyst canister.

1. Remove the four (4) panel screws using a 1/8” Allen wrench

and pull the top panel out. Note the position of the filter so
that you can install the replacement in the same way.


2. Loosen the 1/4” compression nuts on either end and pull the

filter from the tubing.

3. Remove the nuts from the old filter and place them on the

new filter.

4. Install the new filter with the standoff end (visible inside the

housing, supporting the filter element) connected to the
catalyst canister outlet.

5. Firmly tighten the compression nuts on the inlet and outlet

tubing.


4.3 Fuse Replacement

If the unit fails to run when plugged in and turned on, first check the fuse.

1. Disconnect the power cord from the power input module

on the bottom panel of the unit.


2. Using a slotted tip screwdriver, pry up at the slot located

about ¼” below the surface of the power input module,
inside the cord entry socket. This will pull up the fuse
carrier that holds the fuse on the bottom and a spare.

3. Inspect the fuse to see if it is blown. If so, remove it and

replace it with the spare fuse.

4. If the fuse is in good shape, re-install and plug the system
back in.





" If the unit still does not work, unplug the unit again and remove the four socket head

screws located on the corners of the bottom panel. Pull the bottom panel out and inspect

Advertising