A: specifications, Physical description, Operating specifications – Perma Pure MDSS (LLC Integrator Series) User Manual

Page 5: Pecifications

Advertising
background image

Section A: Specifications

5

A:

S

PECIFICATIONS

Physical Description


AG-412 Model

– Vertically surface mounted system in aluminum case with removable cover.



AG-193U Model

– 19” rack mounted system in aluminum case with removable cover. 3U

panel height.

Operating Specifications

Sample Gas Flow Range

0.5-

1.5 Lpm @ 380mm Hg (15” Hg)

Sample gas humidity inlet

Maximum 95% RH at room temperature, non-condensing

Dimensions

305mm x 305mm x 102mm (12” x 12” x 4”)

Weight

3.3 Kg. (7.3 lbs)

Inlet Sample Temperature

1-45°C

Maximum Inlet Pressure

1 Bar (15 psig)

Ambient Temperature Range

1° - 45°C (33° - 113°F)

Inlet Dew Point at rated flow

20°C

Outlet Sample Gas Dew Point
(based on 20C saturated inlet humidity)

-1°C @ 0.5 Lpm sample flow rate
4°C @1.0 Lpm sample flow rate
10°C @ 1.5 Lpm sample flow rate

Inlet Tubing Connection

¼” tube (compression)

Outlet Tubing Connection

¼” tube (compression)

Purge (Pump) outlet Connection

¼” tube (compression)

Power Requirement

230 VAC, 50 Hz, 3A

Ambient operating humidity range

0-95% RH, non-condensing

Advertising