Troubleshooting and Maintenance Guide

Symptom

Diagnostic Fix

No LED(s) and no fan.

AC power input.

Ensure that AC power is
connected.

No LED(s) and fan on.

AC input fuse on control board.

DC output fuse on control board.

Replace fuse as necessary.

Replace fuse as necessary.

LED(s) on and no fan.

AC input fuse on power supply.
+15vdc TB4 on control board.

Replace fuse as necessary.
If not present, replace power

supply.

Impinger remains at

ambient temperature.

Peltier current draw. Should be

above 6 amps when first turned
on.

If unit has two active impingers,

switch Peltier element connections
to other channel

If lower than 6 amps or above 8

amps suspect that Peltier element
has failed.

If problem remains at same

channel, control board defective. If
it changes, Peltier defective

Thermocouple failure LED

is on.

Thermocouple connection TB1, 2.

Ensure proper connection to

control board. If connection good,

replace thermocouple.

Impinger frozen and cooler
indicates ambient

temperature.

Thermocouple placement in heat
exchanger block.

Ensure proper placement. Replace
thermocouple or control board.

Impinger does not reach
set temperature, but is

below ready temperature.

System loading.

Calibration and Set temperature
adjustment.

Ensure system loading is not
exceeding cooler capacity.

Consult with Perma Pure Tech
Support

Impinger temperature

cycles up and down.

Peltier connections on control

board.

Ensure a firm connection on flag

connectors on control board.

Ensure system loading is not

exceeding cooler capacity.

Ready LED does not come
on when impinger is below

7°C.

Ready temperature adjustment.

Consult with Perma Pure Tech
Support.

Water Slip (carryover) in

system.

Impinger temperature. Should be

below 6°C.

Ensure system loading is not

exceeding cooler capacity.
Ambient must not exceed 45°C

(113°F)

Slip LED does not come on

(alarm relay/water slip

option installed).

Water slip sensor connections.

Ensure that all water slip sensor

connections are made.

Clean tip of sensor.
Replace alarm relay/water slip

board as needed.