Sample pump does not

start. Ready and slip

LED(s) are on (alarm

relay/water slip option
installed)
.

Water carryover in system


Ambient air temperature too high


Operation of peristaltic pump(s) or

eductor


Heat sink cooling fan and heat sink

air flow path


Air leaks in sample line

Clean water slip sensor and

restart. Ensure system loading

does not cause water slip


Ambient must not exceed 43°C

(110°F)


Verify that pump or eductor is

draining condensate from

impingers; Replace pump/hoses as
necessary

Fan and air path need to flow

proper amount of air. If debris is
clogging air path or fan is not

functioning properly, unit will loose

efficiency

Verify that there are no air leaks at

any point in the sample line.
Ambient air entry either on the

pressure or vacuum side of the

sample pump will cause excess

condensate and could lead to
water slip.

Sample pump does not
start.

Ready LED is on;

Slip LED is Off

(alarm relay/water slip
option installed)
.

Electrical connections to sample
pump


Verify that proper voltage is
reaching pump.

If voltage is good, replace pump

If voltage is not present, check for
voltage at relay board; if not

present consult with Perma Pure

Tech Support.
If voltage present trace out wiring

issue.

Sample pump starts. No

sample stream.
Ready LED is on;

Slip LED is ON (green)

(alarm relay/water slip

option installed).

Sample line connections


Verify that flow meter needle is

not seated


Filter element


Sample line clogged or deformed


Sample pump suction/ pressure

Verify proper connections.


Adjust as necessary


Check for clogged filter. Replace

element if necessary.


Unclog, reform/replace as

necessary


Sample pump valve

discs/gaskets/diaphragm have

reached the end of service life.
Replace as necessary. Consult with

Perma Pure Tech Support