Section 2: INSTALLATION

2.10 Introduction

This manual has been compiled not only for the care and maintenance of the Welch Dry Vacuum System

now in your possession, but as a helpful reference and guide to prevent many problems which can occur

if used improperly.

2.11 Unpacking

Carefully remove the Dry Vacuum System from the shipping carton. Preserve all paperwork for future

reference. If damage has occurred from shipment a claim must be filed with the carrier immediately;

preserve the shipping carton for inspection by the carrier. If you are required to communicate with your

dealer or Welch Vacuum be sure to include your order numbers for quick identification. Do not return

the system to the factory without first calling for a returned goods number. Call (847) 676-8800

2.12 System Mounting

Rubber feet are attached to the pump. Rubber feet are excellent for applications involving a

semi-flexible surface such as a bench top; they help to isolate noise and eliminate creeping.

2.13 System Location and Enviromental Condition

The Dry Vacuum System should be located preferably in a clean, dry and well ventilated area. Please

be sure not to block the ventilation ports located on the metal shell. The pump should be placed where

the surrounding temperature remains between 10°C and 40°C (50°F and 104°F). Always check to insure

the location chosen is protected from direct or indirect moisture contact. The pump should be located as

closely to its system in order to utilize it most efficiently. The pump is rated for indoor use only.

Maximum altitude 2000 meters. Operating temperature range 5°C to 40°C. Maximum relative humidity

of 80% for temp. up to 31°C decreasing to 50% at 40°C. Rated for ±10% of supply voltage.

Pollution Degree2, Installation Category II.

2.14 Inlet and Outlet Provisions

A inlet hose barb which accepts ¼” I.D. hose makes it easy to connect the system to your concentrator,

rotary evaporator or gel dryer. Since the vacuum system operates in the viscous flow regime, the small

diameter of the hose will generate minimal conductance loss. For best results, Welch recommends the

length of the tubing between the pump and the apparatus be kept as short as possible. Hose clamps

should be used to hold the hose in place.

The hose barb in the outlet port of the pump accepts ¼” ID hose. A vent line can be attached to the

outlet hose barb which allows gases and vapors pumped through the pump to be piped to a fume hood.

Be sure to call Welch technical service prior to start-up at (847) 676-8800 if you have any questions.

WARNING

Don’t operate this pump in an atmosphere

containing flammable or explosive gases or vapors.

WARNING

The motor is thermally protected and will automatically

restart unexpectedly when the overload device resets.

WARNING

Never Block The Outer Port. If the outlet is blocked, pressure will

buildup in the pump which can lead to the pump head bursting

creating the potential of serious injury.

Remove plug from outer port.