Welch Vacuum 8925 Direct Drive Rotary Vane Pump User Manual

Page 25

Advertising
background image

25

7.02 Trouble Shooting Guide

Condition

Probable Cause

Recommended Corrective Action

1. Power off.

1. Check switches and fuses.

Pump

2. Coupling damaged or

2. Check coupling and set screw.

will not

coupling set screw loose.

start

3. Room is too cold.

3. Drain and refill pump with warm oil.

4. Pump mechanism is seized.

4. Pump to be repaired.

1. Pump is contaminated.

1. Flush and change pump oil.

Use a foreline trap.

2. Pump is not filled with oil,

2. Add recommended pump oil.

Pump does not or has low oil level.

reach ultimate

3. Pump has wrong oil in it.

3. Flush and refill with recommended pump oil.

pressure

4. Leak in vacuum system.

4. Locate and eliminate leak source.

5. Dirty foreline trap.

5. Clean out cold traps and replace elements

in coaxial and molecular sieve traps.

6. Gas ballast open.

6. Close gas ballast.

1. Intake or exhaust lines

1. Clean and straighten out lines and

are restricted.

check oil level.

Excessively

2. Coupling damaged.

2. Examine coupling / replace.

noisy

3. Slapping noise at ultimate

3. Clean orifice air passage.

pump

pressure is typical for

some pumps.

4. Inside mechanism damaged.

4. Pump to be repaired.

Pump generates 1. Pump overfilled with oil.

1. Drain excess pump oil.

excessive smoke 2. Pump operating continously

2. Use larger capacity pump, or

or oil mist from above its maximum

modify your vacuum system.

exhaust port

operation pressure.
1. Pump oil contaminated by

1. Flush and change pump oil. Use foreline

process gases, or other

trap in futher. Consider using a oil filtration

Pump oil is

foreign material ingested

system or inert pump oils.

dark, has an

by pump.

unusual color, 2. Pump oil has degraded.

2. Pump has run too low on oil. The

or is dirty

recommended oil was not used. Pump

is running continously above its maximum

operation pressure.

1. Pump is running too cold.

1. Allow pump to run until it warms up to

its operating temperature.

2. Exhaust or intake line

2. Install larger inner diameter tubing.

Pump does not its too narrow.

achieve its rated 3. Pump oil is contaminated.

3. Flush and change pump oil.

pumping speed

Use a foreline trap in futher.

4. Very dirty trap or

4. Clean out cold traps and replace elements

intake line.

in coaxial and molecular sieve traps.

Clean or replace vacuum piping.

Motor is

1. Voltage selection

1. Convert motor to be compatible with

excessively noisy is incorrect.

electrical supply. See Section:

4.03 / 4.04 Changing voltage setting.

Advertising