Plug the sensor probe into 33% salt
bottle. Hold down and under
normal mode to enter 33% calibration
(Fig.17). “CAL” and calibrating value
(32.7% if at

are blinking on the

LCD with current temperature at the left.

Meter is now calibrating, and will finish
in about 60 minutes when “CAL” and
humidity stop blinking. (Fig.18)

o

25 C)

The meter defaults to be calibrated the
humidity with 33% and 75% salt solution.
The ambient condition is recommended

o

to be at 25 C and stable humidity(better
to be close to the calibrating value). To
abort calibration, hold down for more
than 1second at any time.

(7722/77232)

RH CALIBRATION

%

%

O

O

Fig. 17

%

%

O

O

Fig. 18

After 33% calibration, plug the sensor
probe into 75% salt bottle, then press
to enter 75% calibration (Fig.19).

CAUTION:
Do not calibrate the humidity without
the default calibration salt. Otherwise,
it will cause permanent damage.
Contact the dealer for calibration salt
or services.

M /

X

N

AVG

CAL

ESC

CAL

ESC

SET

SET

RESET

RESET

33% calibration

75% calibration