PC CONNECTION

The meter can do PC link for

data analysis

on-line

logging and

via RS232

interface and software.
The protocol is as follows.

B.Format (

ASCII)

Model 7721/77231
Cxxxxppm:Txxx.xC(F) LRC CRLF

Description: $CO

2

:Air LRC CRLF

Model 7722/77232
Cxxxxppm:Txxx.xC(F):Hxx.x%:
dxxx.xC(F):wxxx.xC(F) LRC CRLF

Description: $CO :Air:RH:DP:WBT LRC

2

CRLF

A.9600 bps, 8 data bits, no parity.

SPECIFICATION

7721

7722

77231

77232

Resolution

Accuracy

Pressure
Dependence

Range
Resolution
Accuracy

Range

N/A

0.0~99.9%

N/A

0.0~99.9%

Resolution

N/A

0.1%

N/A

0.1%

±3%(10~90%)

±3%(10~90%)

±5%(others)

±5%(others)

Warm up
Operating
Storage
Power

±30ppm±5%rdg

(0~5000)

Not specified for out of scale

-10.0~60.0℃ (14~140℉)

0.1℃/0.1℉

+1.6% reading per kPa deviation from

normal pressure, 100kPa

Temp.

CO

2

Humidity

Range

1 ppm

1 ppm

0~2000ppm

2001~9999(

out of scale

)

0~5000ppm

5001~9999(

out of scale

)

±50ppm±5%rdg

(0~2000)

Not specified for out of scale

Accuracy

30 seconds

0~50℃, 0~95%RH

(avoid condensation)

±0.6℃/ ±0.9℉

-20~60℃, 0~99%RH

(avoid condensation)

12V adaptor

N/A

N/A