Ronan X96S MASS FLOW GAUGE User Manual

Page 59

Advertising
background image

55

The Density Calibration is complete.

For future reference, document these items:

(a) Environmental/process conditions (densities, temperatures, etc.) that influence the reference/ calibration.
The next time a calibration is performed, you will need to duplicate the conditions, or account for the
differences.

(b) All changes made to factory-default settings such as time constant, reference constants (loading factor),
etc.

(c) A record of "counts" being received from the detector may assist with future troubleshooting efforts.