Setting or verifying flow rate - low flow – SKC 224-52 AirChek 52 User Manual

Page 12

Advertising
background image

10

Setting or Verifying Flow Rate - Low Flow

Ensure pump has run for fi ve minutes before calibrating. Set the
pump fl ow rate at approximately 1.5 L/min (see Setting or Verifying
Flow Rate for High Flow Applications on page 7
).

Connect a constant pressure controller (CPC) (Figure 2, #3) to the
pump intake (Figure 2, #1) using 1/4-inch Tygon tubing.

Using 1/4-inch tubing, connect the intake of the CPC to the outlet
of a single adjustable low fl ow holder (Figure 2, #4).

Insert an opened sorbent tube into the
holder’s rubber sleeve with the arrow on
the tube pointing toward the holder.

Connect a calibrator to the exposed end
of the sorbent tube.

Loosen the brass fl ow adjust screw on the low fl ow holder. Activate
the pump by pressing the on/off button (Figure 1, #3).

Adjust the fl ow rate by turning the fl ow
adjust screw (Figure 2, #6) on the adjustable
low fl ow holder until the calibrator indicates
the desired fl ow.

Do not adjust the fl ow on the pump.

Adjust the fl ow only by using the fl ow
adjust screw on the low fl ow holder.

Figure 2

Single Adjustable Low Flow Holder attached to AirChek 52

1 Pump

intake

2 Tygon

tubing

3 Constant

Pressure

Controller (CPC)

4 Single adjustable low

fl ow holder

5 Protective cover for sample

tube

6 Flow adjust screw

(needle valve)

7 Collar

clip

Calibration train with

tube in low fl ow holder

Flow adjust

screw

Turn screw to

adjust fl ow.