Maintenance, Pump inlet filter, Replacing the battery pack – SKC 224-52 AirChek 52 User Manual

Page 18

Advertising
background image

16

Maintenance

Pump Inlet Filter

The AirChek 52 Sampler is fi tted with a fi lter/trap just below the
pump intake. This prevents particles from being drawn into the pump
mechanism. The fi lter should be replaced periodically. Follow this
procedure:

1. Grip the inlet hose barb (Figure 1, #4 on page 9)

and unscrew it counterclockwise.

2. Holding the pump upside down, use a small

fl athead screwdriver or tweezers to remove
fi lter.

3. Place a new fi lter in the intake. Ensure the

rubber gasket inside the inlet hose barb is still
in place.

4. Screw the inlet hose barb back into place.

Replacing the Battery Pack

To enhance battery life, SKC ships battery packs separate from
the pump (
see Charging the Battery Pack on page 5). Once installed,
completely charge the battery pack before operating the pump. It
may be necessary to charge the battery a few times before maximum
battery capacity is achieved.

1. Loosen two security screws on bottom of pump case and allow them to

fall to the out position.

2. Pinch battery pack release arms and pull

the bottom half of the pump case away
from the top half.

3. Align arms on top half of case with slots

on either side of the new battery pack
housing. Ensure the back of the battery
pack with the charging jack is on the
same side as the belt clip.

4. Slide battery pack up until arms click into

place.

5. Push each security screw into the battery pack housing and tighten.

Use of a repaired or rebuilt battery pack voids any warranty
and the UL Listing for intrinsic safety.

For information on SKC pump battery packs,

visit http://www.skcinc.com/instructions/1756.pdf.

Batt ery
pack
release
arms

Batt ery
pack

Security
screws
in out
position

Advertising