Calibration with a QuickTake 30 Sample Pump

1. Ensure the BioStage is fully assembled with a representative agar plate in

place.

2. Connect the rotameter outlet to one end of the ¼-inch ID tubing.

3. Insert the small end of a Luer adapter into the free end of ¼-inch tubing.

4. Insert the large end of the Luer adapter into the short length of the

3

/

8

-inch

adapter tubing.

5. Install the free end of the adapter tubing onto the calibration adapter.

6. Press the calibration adapter onto the BioStage inlet until a fi rm seal is established.

7. Connect the BioStage outlet to one end of the ¼-inch ID tubing.

8. Insert the small end of a second Luer adapter into the free end of the ¼-inch

tubing.

9. Insert the large end of the second Luer adapter into one end of the

3

/

8

-inch

sampling tubing and the pump inlet adapter into the other end.

10. Insert the pump inlet adapter into the pump inlet.

11. Calibrate the sample pump to 28.3 L/min for a Standard BioStage or

14.15 L/min for BioStage 200.

The calibration adapter and rotameter are for use with both models of BioStage. See Accessories.