Teledyne LeCroy WaveRunner Automation Command User Manual

Page 18

Advertising
background image

Automation Command and Query Reference Manual - Control Reference

Sets/Queries the amplitude of the signal on the AUX OUT connector. Note that this is the amplitude of
the signal into a 1Mohm load. Into 50 ohms the output voltage will be halved (since the source
impedance is nominally 50 ohms). Units are Volts.

Description

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Set the amplitude of the signal from the AUX OUT connector
' to 0.6 V into 1 megohm, or 0.3 V into 50 ohms.
app.Acquisition.AuxOutput.Amplitude = 0.6

Example

AuxInCoupling

Enum

Sets the input coupling for the Auxiliary input path.

Description

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Set the coupling of the Auxiliary socket, when used as an input, to
ground.
' In this condition, no input signal reaches the instrument.
app.Acquisition.AuxOutput.AuxInCoupling = "GND"

Example

Values

DC50

DC, 50ohms coupling

GND

Grounded

AuxMode

Enum

Configures AUX Output type in WR and WS series of scopes.

Description

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Set the Auxiliary output to trigger output signal.
app.Acquisition.AuxOutput.AuxMode = "TriggerOut"

Example

Values

Off

No output

PassFail

Pulse on Pass fail condition

TriggerEnabled

Trigger enabled signal from trigger circuitry.

TriggerOut

Internal trigger output signal from trigger circuitry

1-10

918500 RevA