App.measure.px.operator (paramengine = "tie"), Hysteresis, Range – Teledyne LeCroy WaveRunner Automation Command User Manual

Page 449: Annotate

Advertising
background image

Automation Command and Query Reference Manual - Processor Reference

Hysteresis

Double

From 0 to 10 step 0.1

Range

Hysteresis

Double

TIE

app.Measure.Px.Operator (ParamEngine = "TIE")

TIE is "Time Interval Error", or the error in expected arrival time of trnasotions in either a data stream or a clock signal.
It is the heart of most jitter measurments (where only one signal is under analysis). The Skew processor is closely
related to this function in cases where measurements are using a separate reference clock.

WARNING: The TIE processor is appropriate for analysis of "real-time" acquired waveforms for jitter and timing
variations. It will give incorrect results for equivalent-time type waveforms.

Annotate

Integer

From 1 to 10 step 1

Range

Annotate

Integer

BaseFrequency

Double

DataIsNRZ

Bool

FindBaseFrequency

Action

FindLevel

Action

FrequencyMultiplier

Double

Hysteresis

Double

IncludeVirtualEdges

Bool

IntervalsEdgeEdge

Integer

IntervalType

Enum

LevelType

Enum

MaxComboIntervals

Integer

PercentLevel

Double

PermitGTHalfUI

Bool

PLL1TransportDelay

Double

PLLCompensateForMissingEdges

Bool

PLLFrequency

Double

PLLType

Enum

SignalType

Enum

Slope

Enum

Summary

String

UseAllEdges

Bool

UseBaseFrequency

Enum

UseMultiEdgeCombos

Bool

UsePLL

Bool

VirtEdgeType

Enum

2-118

918500 RevA

Advertising