TIF IR-1 Infrared Refrigerant Leak Detector User Manual

Page 7

Advertising
background image

filter Replacement

A filter, located in the tip of the wand, blocks moisture and other

contaminants from the sensor. When this filter becomes wet, it

restricts the flow of air and must be replaced.
To replace the filter:

1. Unscrew and remove the tip of the sensor to expose the white

filter.

2. Remove and discard the filter.

3. Position a new filter in the tip so the round end of the filter is

closest to the tip of the wand.

NOTE: Use only TIF supplied replacement filters.

Storage

The battery should have a 40%-50% charge during prolonged

storage of a month or longer. The correct storage environment is

critical to battery life.
Storage environment: <80% RH for detector and battery.
For 80% battery recovery:

-4ºF (-20°C) to 140ºF (60°C) less than 1 month

-4ºF (-20°C) to 113ºF (45°C) less than 3 months

-4ºF (-20°C) to 68ºF (20°C) less than 1 year

NOTE: Battery life will be reduced significantly if the battery is

stored with a full charge and/or at high temperatures.

Accessories and Replacement Parts

Optional Accessories

Order No.

Extension Probe (9 inch) ......... TIFIR-4

Rigid Extension Probe ............. TIFIR-5

Replacement Parts

Order No.

100-240 VAC Charger ............. TIFIR-6

12 VDC Charger ...................... TIFIR-7

Filters (5) and O-rings (3) ........ TIFIR-8

Lithium-ion Battery .................. TIFIR-9