Appendix b troubleshooting, Table b-1 battery troubleshooting procedures, Symptom – ZOLL AutoPulse Rev A User Manual

Page 39: Possible cause, Recommended action

Advertising
background image

P/N. 12457-001 Rev. 2

Page B-1

Power System User Guide

Appendix B Troubleshooting

This Appendix details symptoms, possible causes, and recommended actions for difficulties you might
have with your AutoPulse Power System. Table B-1 provides troubleshooting procedures for the Battery.
Table B-2 provides troubleshooting procedures for the Multi-Chemistry Battery Charger.

Table B-1 Battery Troubleshooting Procedures

Symptom

Possible Cause

Recommended Action

Battery Status Check
LEDs do not illuminate.

The Battery’s status is
unknown.

Place the Battery in one of the charging bays of the
Multi-Chemistry Battery Charger.

1.

If its Charging LED is illuminated, the Multi-
Chemistry Battery Charger is attempting to
restore the Battery. Refer to Chapter 4,
“Operating the AutoPulse Multi-Chemistry
Battery Charger”
for more information.

2.

If the Charger’s Fail LED is illuminated, the
Battery has failed. Replace the Battery. Refer to
Section 6.2.4, “Reaching the End of AutoPulse
Li-Ion Battery Service Life” or Section 6.2.5,
“Reaching the End of AutoPulse NiMH Battery
Service Life” for
more information.

Battery will not fully insert
into the Multi-Chemistry
Battery Charger.

Protective plastic cap
was left on the Battery.

The Battery may be
damaged.

The Battery Charger’s
charging bay might be
obstructed.

Remove protective cap.

Inspect the guide rails around the Connector for
damage. If the guides are damaged, replace the
Battery.

Inspect the Battery Connector for damage. If the
Connector is damaged, replace the Battery.

Unplug the alternating current (AC) from the
Multi-Chemistry Battery Charger. Check the
charging bay to ensure that no debris has
accumulated in the bay.

The Battery will not fully
insert into the AutoPulse
Platform.

Protective plastic cap
was left on the Battery.

The Battery may be
damaged.

The AutoPulse
Platform’s Battery
compartment might be
obstructed.

Remove protective cap.

Inspect the guide rails around the Connector for
damage. If the guides are damaged, replace the
Battery.

Inspect the Battery Connector for damage. If the
Connector is damaged, replace the Battery.

Check the AutoPulse Platform’s battery
compartment to ensure that no debris has
accumulated in the bay.

Advertising