Manual Defibrillator Testing

9650-0904-01 Rev. K

ZOLL R Series Operator’s Guide

13–7

10 mm ±1.0 mm high

Code Readiness Log

Each automatic and manual defibrillation test result is stored in internal non-volatile memory,
the Code Log. A total of 1000 Code Readiness test records can be stored in internal memory.
When the Code Readiness Log is full, the oldest records are erased on a first-in-first-out (FIFO)
basis.

The Code Readiness Log can be transferred to an external computing device (see “Transferring
Files to an External Device” on page 12-1).

If configured, the R Series prints a Code Readiness Test Report following the completion of
each automatic defibrillator test.

To print the Code Readiness Log:

1. Turn the Mode Selector to ON.

2.

Press the Manual Mode softkey.

Note:

If the CONFIRM MANUAL MODE message displays, press the Confirm softkey to
enter the Manual mode of operation.

3.

Press the Report Data softkey.

4.

Press the Test Log softkey.

A menu appears with the print options.