ELAPSED TIME

When enabled, this feature indicates the elapsed time since the unit was first turned on. It is
displayed in the lower left corner. The elapsed time is displayed in MM:SS format up to 99:59.
If the unit is on for over 100 minutes, the elapsed time will roll over to 0. The elapsed time will
be maintained for up to 10 seconds after power down. This will give the operator adequate time
to change the unit’s battery without resetting the elapsed time.

FULLY RELEASE

If configured to do so, this message reminds rescuers to lift (fully release) their hands from the
patient’s chest during compressions to allow full recoil. By default, this feature is not enabled.

MONITOR

The MONITOR message appears when the ECG cable is connected to its input connector and
attached to the patient, but the therapy electrodes are not connected. The unit selects lead II and
sets the ECG size automatically (you cannot change the Lead and ECG size).

Switching to Manual Mode Operation

Press the Manual Mode softkey on the front panel of the unit to enter the Manual mode of
operation.

The message CONFIRM MANUAL MODE and the Confirm softkey will be displayed. Press
the Confirm softkey to enter the Manual mode of operation. If the Confirm softkey is not
pressed within 10 seconds, the unit will revert back to AED operation.

Note:

If the unit has been configured not to confirm the transition from AED mode to
Manual mode, the unit will immediately enter Manual mode.

120J SEL.

Manual

Mode

00:01

ECG

Analyze

120J SEL.

CONFIRM MANUAL MODE

Confirm

ECG

00:01