Programming

18

Step

Comment

CAL 12

High Alarm or Pulse Output Factor - whole numbers.

CAL 13

High Alarm or Pulse Output Factor - digits after the decimal
point.

CAL 12 & 13 program the flow rate above which the high
alarm relay will close. The value can be programmed in the
range 0 to 999,999.

If the scaled pulse output is selected (see CAL0 = 1), then the
value will represent the total per pulse, eg. 5 litres per pulse.

CAL 14

4mA Retransmission - whole numbers.

CAL 15

4mA Retransmission - digits after the decimal point.

CAL 14 to CAL 15 represent the flow rate at which 4mA will
be output. If the 4-20mA retransmission is not installed, these
parameters can be ignored.

CAL 16

20mA Retransmission - whole numbers.

CAL 17

20mA Retransmission - digits after the decimal point.

CAL 16 & 17 are the flow rates at which the output will reach
20mA.

SOFT

Software Version.