background image

IM 11M13A01-04E

8-20

8.7.3.2Procedure

Use the parameter code table below to set fuel values.

Table 8.17 Setting Fuel Values

Set item

Parameter code

Set value Engineering units

Amount of water

F20

0 to 5

m

3

/kg (m

3

)

vapor in exhaust gas
Theoretical

F21

1 to 20

m

3

/kg (m

3

)

amount of air
X value

F22

0 to 19.99

Absolute humidity

F23

0 to 1

kg/kg

of the atmosphere

T8.17E.EPS

8.7.3.3Default Values

When the analyzer is delivered, or if data are initialized, parameter settings are by
default, as shown in Table 8.18.

Table 8.18 Default Settings of Fuel Values

Item

Default setting

Amount of water vapor in exhaust gas

1.00 m3/kg (m3)

Theoretical amount of air

1.00 m3/kg (m3)

X value

1.00

Absolute humidity of the atmosphere

0.1000 kg/kg

T8.18E.ESP

8.7.4

Setting Purging

Purging is to remove condensed water in the calibration gas pipe by supplying a span
calibration gas for a given length of time before warm-up of the detector. This prevents
cell breakage during calibration due to condensed water in the pipe.

Open the solenoid valve for the automatic calibration span gas during purging and after
the purge time has elapsed, close the valve to start warm-up.

Purging is enabled when the cell temperature is 100

Њ C or below upon power up and the

purge time is set in the range of 1 to 60 minutes.

F8.05.EPS

Displayed alternately

Figure 8.5 Display during Purging

8.7.4.1Procedure

Use the parameter-code table below to set the purging time.

The allowable input ranges from 0 to 60 minutes.

Table 8.19 Prameter Code

Set item

Parameter code Set range Units

T8.19E.EPS

Purging time

F15

0 to 60

minutes

8.7.4.2Default Value

When the analyzer is delivered, or if data are initialized, purging time is set to 0
minutes.