2. Specifications

Contact Related Items: Selection of contact input 1 and 2, selection of contact output 1

and 2 (abnormal, high-high alarm, high-alarm, low-alarm, low-low alarm,

maintenance, calibrating, range switching (switched range is fixed), warmingup,

calibration-gas pressure decrease, flameout gas detection (answerback of contact

input)

Converter Output: One mA analog output point (4 to 20 mA DC (maximum load

resistance of 550

⍀)) with mA digital output point (HART) (minimum load

resistance of 250

⍀).

Range: any setting between 0 to 5 through 0 to 100 vol% O

2

in increments of 1 vol%

O

2

, and partial range is available (Maximum range value/minimum range value

1.3 or more)

For the log output, the minimum range value is fixed at 0.1 vol% O

2

.

4 to 20 mA DC linear or log can be selected.

Input/output isolation

Output damping: 0 to 255 seconds.

Hold/non-hold selection, preset value setting possible with hold.

Contact Output: Two points, contact capacity 30V DC 3A, 250V AC 3A (resistive load)

One of the output points can be selected to ether normally energized or nor

mally de-energized status.

Delayed functions (0 to 255 seconds) and hysteresis function (0 to 9.9 vol% O

2

)

can be added to high/low-alarms.

The following functions are programmable for contact outputs.

(1) Abnormal, (2) High-high alarm, (3) High-alarm, (4) Low-low alarm,
(5)Low-alarm, (6) Maintenance, (7) Calibration, (8) Range switching answer-
back, (9) Warm-up, (10) Calibration-gas pressure decrease (answerback of
contactinput), (11) Flameout gas detection (answerback of contact input).

Contact Input: Two points, voltage-free contacts

The following functions are programmable for contact inputs:

(1) Calibration-gas pressure decrease alarm, (2) Range switching (switchedrange
is fixed), (3) External calibration start, (4) Process alarm (if this signal is
received, the heater power turns off)

Self-diagnosis: Abnormal cell, abnormal cell temperature (low/high), abnormal

calibration, A/D converter abnormal, digital circuit abnormal

Calibration: Method; zero/span calibration

Calibration mode; automatic, semi-automatic and manual (All are operated using
optical switches). Either zero or span can be skipped.

Zero-calibration gas concentration setting range: 0.3 to 100 vol% O

2

(in 0.01

vol% O

2

). Span-calibration gas concentration setting range: 4.5 to 100 vol% O

2

(in 0.01 vol% O

2

).

Use nitrogen-balanced mixed gas containing 10 vol% O

2

scale of oxygen for

standard zero-gas, and 80 to 100 vol% O

2

scale of oxygen for standard span gas.

Calibration period; date/time setting: maximum 255 days/23 hours.