6-3 transducer block, 6-3-1 transducer block ph202, 6-3 transducer block -32 – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 37: 6-3-1 transducer block ph202 -32, 6�-3 transducer block

Advertising
background image

IM 12A00A01-61E

3-32 Foundation Fieldbus

3-6�-3-1 Transducer Block PH202

Index Parameter name

Factory

Valid

Description

Default

Range

2000 BLOCK HEADER

TAG: “TB”

General information about the function block

2001 ST_REV

-

-

The revision level of the static data associated with the function

block. The revision value will be incremented each time a static

parameter value in the block is changed.

2002 TAG_DESC

“ “

The user description of the intended application of the block.

2003 STRATEGY

1

The strategy field can be used to identify grouping of blocks. This

data is not checked or processed by the block.

2004 ALERT_KEY

1

The identification number of the plant unit. This information may be

used in the host for sorting alarms, etc.

2005 MODE_BLK

AUTO

The actual, target, permitted, and normal modes of the block

2006 BLOCK_ERR

-

This parameter reflects the error status associated with a block. It

is a bit-string, so that multiple errors can be shown.

2007 UPDATE_EVT

-

The alert is generated by any change to the static data.

2008 BLOCK_ALM

-

The block alarm is used for all configuration error, hardware

connection failure or system problems in the block. The cause of

the alert is entered in the subcode field. The first alert to become

active will set Active status in Status attribute.

2009 TRANSDUCER_DIRECTORY

A directory that specifies the number and starting indices of the

transducers.

2010 TRANSDUCER_TYPE

Standard

PH, ORP transducer block.

pH/ORP

2011 XD_ERROR

-

The error code in transducer:

- No failure,

- Electronics failure,

- I/O failure,

- Mechanical failure

2012 COLLECTION_DIRECTORY -

A directory that specifies the number, starting indices and DD item

IDs of the data collection in each transducer within a transducer

block.

2013 PRIMARY_VALUE_TYPE

PH

pH, None

Type of measurement represented by primary value.

2014 PRIMARY_VALUE

-

-2 to 16

Primary value of the instrument is pH.

2015 PRIMARY_VALUE_RANGE

-2.00 to 16.00 -2 to 16

The pH range of the instrument. (can not change)

2016 SENSOR_TYPE_PH

pH/ORP

pH/ORP

PH is measured with a glass- and a reference electrode.

sensor

sensor

Redox is measured with a metal- and a reference electrode. When

a glass, metal and ref. Electrode are combined one can measure

pH and Redox simultaneously.

2017 SENSOR_MV

-

-

The output of the electrode in mV

2018 CAL_POINT_HI

16

-2 to 16

Highest calibration point

2019 CAL_POINT_LO

-2

-2 to 16

Lowest calibration point

2020 CAL_MIN_SPAN

1

1 to 18

Minimum span between two calibration points

2021 SLOPE

100

70 to 110%

Sensitivity of the glass electrode

2022 SLOPE_UNIT

%

%

Slope is presented as percentage of theoretical slope

(59,16 mV/pH equals 100%)

2023 ZERO

0

-120 to 120 mV,

A balanced electrode system gives 0 mV output at pH=7.

mV -2 to 16 pH

Zero (asymmetry potential) indicates the offset in mV.

2024 ZERO_UNIT

mV

mV, pH

As an alternative to Asymmetry Potential, the Zero point can be

used to define and calibrate the EXA pH transmitter conform to the

DIN standard for instruments No. IEC 746-2.

2025 ISOPOTENTIAL_PH

7

-2 to 16 pH

This is the pH value at which the sensor has the same mV output

independent of the process temperature.

2026 SENSOR_CAL_METHOD

0

1 point,

not used

2 point

2027 SENSOR_CAL_DATE

-

till 2104

Date the sensor was last calibrated.

2028 SECONDARY_VALUE

-

-30 to 140°C ,

Temperature value

-20 to 280 °F

3-6�-3 Transducer Block