4-3-4. function block parameters do202, 4-3-4. function block parameters do202 -17 – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 75

Advertising
background image

Profibus 4-17

IM 12A00A01-61E

4-4-3-4. Function Block Parameters dO202

Parameter

default

Alternatives

Unit r/W Slot, Index data Type

(byte,bit) (bytes)

Primary value parameters

Primary value Type

dissolved oxyg. (65520)

R

4,27 unsigned16 (2)

Measured Value

ppm R

4,28

DS-33 (5)

Status

R

4,28

DS-33 (5)

Sensor type DO

polarographic (2)

galvanic (1)

R/W 4,31

unsigned16 (2)

Primary value unit

ppm (1423)

% saturation (1342), ppb (1424)

R

4,30

unsigned16 (2)

Secondary value Parameters

Value

°C

R

4,43

DS-33 (5)

Status

R

4,43

DS-33 (5)

Secondary value unit

°C (1001)

°F (1002)

R/W 4,44

unsigned16 (2)

Sensor temp comp

auto (2)

manual (1)

R/W 4,45

unsigned8 (1)

Sensor temp man value

25

°C

R/W 4,46

float (4)

Temperature sensor

NTC22k (160)

Pt1000 (148), Pb36 (152)

R/W 4,47

unsigned16 (2)

Other value Parameters

Sensor current

nA

R

4,49

float (4)

%Saturation

%

R

4,50

float (4)

Calibration Parameters

Zero calibration

disabled (0)

enabled (1)

R/W 4,97 (1.0) unsigned32 (4)

Zero current

0.0

-199.9 ~ 199.9

nA

R/W 4,33

float (4)

Zero current limit

199.9

nA

R/W 4,51

float (4)

Sensitivity

7.5

1.0 ~ 1999.9

nA/

ppm R/W 4,34

float (4)

Stable time

60

5 ~ 600

s

R/W 4,35

float (4)

Stable value (sensitivity)

0.1

0 ~ 50.0

ppm R/W 4,36

float (4)

Stable value (zero)

0.1

0 ~ 50.0

ppm R/W 4,37

float (4)

Calibration pressure

101.3 (kPa)

0 ~ 999

kPa R/W 4,39

float (4)

Compensation Parameters

Salinity compensation

disabled (0)

enabled (1)

R/W 4,97 (1.4) unsigned32 (4)

Salinity

0.0

0 ~ 99.9

ppt

R/W 4,38

float (4)

%Saturation pressure

101.3 (kPa)

0 ~ 999

kPa R/W 4,41

float (4)

diagnostic Settings

Calibration not stable (E1) Hard fail (1)

Soft fail (0)

R/W 4,96 (1.0) unsigned32 (4)

Zero out of limits (E2)

Hard fail (1)

Soft fail (0)

R/W 4,96 (1.1) unsigned32 (4)

Temp. sensor open (E7) Hard fail (1)

Soft fail (0)

R/W 4,96 (1.6) unsigned32 (4)

Temp. sensor shorted (E8)

Hard fail (1)

Soft fail (0)

R/W 4,96 (1.7) unsigned32 (4)

Primary value exceeds

limits (E9)

Hard fail (1)

Soft fail (0)

R/W 4,96 (2.0) unsigned32 (4)

Calibr. timer expired (E16) Soft fail (0)

Hard fail (1)

R/W 4,96 (2.7) unsigned32 (4)

Passcode Configuration

Passcode Maintenance

000

111,333,777,888,123,957,331,546,847

R/W 4,62

float (4)

Passcode Commisioning 000

111,333,777,888,123,957,331,546,848

R/W 4,63

float (4)

Passcode Service

000

111,333,777,888,123,957,331,546,849

R/W 4,64

float (4)

display Settings

Manual pressure in

maintenance menu

disabled (0)

enabled (1)

R/W 4,97 (1.7) unsigned32 (4)

Auto return

enabled (1)

disabled (0)

R/W 4,97 (1.3) unsigned32 (4)

Logbook Configuration

Power up

logbook2 (3)

not logged (1), logbook1 (2)

R/W 4,69

unsigned8 (1)

Power down

logbook2 (3)

not logged (1), logbook1 (2)

R/W 4,70

unsigned8 (1)

System error

logbook1 (2)

-

R

4,71

unsigned8 (1)

Defaults loaded

not logged (1)

logbook1 (2), logbook2 (3)

R/W 4,72

unsigned8 (1)

Logbook erased

not logged (1)

logbook1 (2), logbook2 (3)

R/W 4,73

unsigned8 (1)

Init performed

logbook1 (2)

-

R

4,74

unsigned8 (1)

Error on

not logged (1)

logbook1 (2), logbook2 (3)

R/W 4,75

unsigned8 (1)

Error off

not logged (1)

logbook1 (2), logbook2 (3)

R/W 4,76

unsigned8 (1)

Temperature adjust

logbook1 (2)

not logged (1), logbook2 (3)

R/W 4,77

unsigned8 (1)

Advertising