5-3-6. Communication setup

Code 60

*COMM.

The settings should be adjusted to suit the communicating device connected

to the output. The communication can be set to HART

®

or to PH 201*B

distribution (see Appendix 2).

*ADDR.

Select address 00 for point to point communication with 4-20mA transmission.

Addresses 01 to 15 are used in multi-drop configuration (fixed 4mA output).

Code 61

*HOUR

The clock/calendar for the logbook is set for current date and time as reference.

*MINUT

*SECND

*YEAR

*MONTH

*DAY

Code 62

*ERASE

Erase logbook function to clear the recorded data for a fresh start. This may be

desirable when re-commissioning an instrument that has been out of service

for a while.

5-3-7. General

Code 70

*LOAD

The load defaults code allows the instrument to be returned to the default set

up with a single operation. This can be useful when wanting to change from

one application to another.