4. quick reference for parameter setting – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 77

Advertising
background image

IM 12D06A03-01E

Appendix 2-5

12-4. Quick reference for parameter setting

Preparation for operation entails deciding maintenance-mode settings and measurement range set-

ting-related issues. Service level (mode) settings are described in Sec. 5-3. Shipping-time measure-

ment parameter setting defaults are shown in Table 12-1. In general there is no need to change these

defaults.

Table 12-1 Default values of parameters.

T6.1E.eps

25°C

Linear

0 to 1000mS

0 to 100%

20

31

35

1.88

ON(1)

-

OFF(0)

OFF

NaCl

0

-

0.0.0

3

50

12

32

55

21

22

23-28

52

70

Code

Cell constant

Auto return functions

Calibrate temperature

"Burnout" on fail

Display in Weight%

Temp. coefficient

Matrix temp. compensation

Compensation matrix

Passwords

Revert to factory defaults

Corresponding Parameter

Default

When you should change

Refer to

Reference temperature

Output characteristic

Output table

To use different reference temp.

Sec. 5-3-3

For non-linear characteristic

Sec. 5-3-4

For non-linear characteristic

Sec. 5-3-4

When detector at least 30mm from metal pipe walls

Sec. 5-3-1

Usually corrected by calibration. Seldom

nevessary to enter value by service code 37

To lock HOLD

Sec. 5-3-5

To calibrate against more accurate thermometer Sec. 5-3-2

(esp. when using one-point calibration)

Can set output "burnout" upscale/downscale

Sec. 5-3-4

if abnormal

To display concentration units

Sec. 5-3-5

Temp. coefficient other than NaCl

Complex compensation

Sec. 5.3.3

Can be user defined

To protect settings

Sec. 5-3-5

To revert to factory defaults

Sec. 5-3-7