7 system configuration output format – Yokogawa DA100 User Manual

Page 72

Advertising
background image

IM DA100-11E

7-7

Output Format

7

7.7

System Configuration Output Format

The measurement interval and system connection data are output in the following format by the TS5
+ “Device Trigger (GET)” + CF command:

M : sssssCrLf
S1 : 0=MMMMMM(DD)1=MMMMMM(DD)~5=MMMMMM(DD)CrLf

Slot number

Each symbol denotes the following:
M:

Measurement interval mark

sssss:

Measurement interval; output down to one decimal place (Example: 10.0
for a measurement interval of 10 sec.). The unit is “second.”

S1:

Subunit number

I ------------------ Main unit (DA100 extended type:DA100-2)
0 ------------------ Subunit or DA100-1 (DA100 stand-alone model)
1 to 5 ------------ Subunit
E ----------------- End mark

MMMMMM:

Module name (6 characters)

COMM ---------- Communication module
RELAY --------- Relay output module
REMOTE ------- Remote module
INPUT ---------- Universal input module
mA --------------- mA-input module
AC --------------- Power monitor module
STRAIN -------- Strain input module
PULS ------------ Pulse input module
DI ---------------- Digital input module
ERROR --------- Module error

(DD):

Internal code (hexadecimal, ASCII, 2 characters)

Note

• The number and data of subunits not connected in the system settings are not output.