Pro-Cut Warthog User Manual

Page 51

Advertising
background image

WWW.PROCUTUSA.COM

PG /

Ø51

With the display in 'H' mode, and the 0.800" / 20.32mm thick calibration block in place, press the CALI-
BRATION button, the same one used to set the LRO Acceptance number. You will hear a beep. When you
have done this, you can press the Reset button to exit the H mode or use the Start button again if you wish
to read the DRO numbers out or switch between inch and mm display mode.

CALIBRATION STEP 3:

G2X CALIBRATION

CONTINUED

The cutting head and slide sensor are now calibrated to read thickness correctly. To check calibration you
can put the gage block between the inserts again and re-tighten. If you read between 0.796" / 20.22mm
and 0.804" / 20.42mm you are properly calibrated.

CALIBRATION STEP 4:

If you rotate the adjustment knobs, you will see the DRO display transition to thickness display. After 8 seconds of inactiv-
ity the display will alternate between the thickness reading and whatever else had been on screen. Operation of the DRO
remains unchanged. You turn on the motor and press the Start button.

USING THE G2X