12 troubleshooting – MAHLE RTI NTO-200 (REV A) User Manual

Page 13

Advertising
background image

12

TROUBLESHOOTING

Fault

Probable Cause

Solution

NTO-200 will only
power up for 5
seconds.

An error occurred during
operation

Power up generator and note error
codes on display. Refer to previous
page for more details

No pressure at inlet

Check the following:
1.) NTO-200 is connected to
compressed air supply.
2.) Compressed air supply pressure
meets required specification.
3.) Hoses for any damage.

Internal leak

Ensure cartridge is correctly installed
in grip and lock ring is secure.
Contact NTO-200 supplier.

Internal blockage

Check cartridge.
Refer to "Reduced flow of nitrogen
from NTO-200.”

No flow of nitrogen
from NTO-200

An error has occurred

Refer to error codes for more details.

Internal blockage

Check cartridge.

Cartridge is incorrectly
installed.

Disconnect NTO-200 from
compressed air supply and check that
cartridge correctly connected.

Gasket damaged or
incorrectly installed.

Remove cartridge and check that
gasket is correctly installed.

Inlet pressure too low

Check the following:
1.) Hoses for any damage.
2.) Check compressor.

Internal leak

Ensure cartridge correctly installed in
grip.
Contact NTO-200 supplier.

Reduced flow of
nitrogen from NTO-
200

Outlet pressure exceeded
50% of inlet pressure.

Check inlet pressure falls within range
specified in Tech Specifications.
Check tire pressure. If pressure too
high, press and hold ON/OFF button
to reduce pressure.

Advertising