2 operating and display elements, 3 connections and interfaces, Fig. 5 – LumaSense Technologies IGA 740 User Manual

Page 14: Operating and display elements

Advertising
background image



Technical Description

10

KLEIBER

740

4.1.2

Operating and display elements

The controls and indicators are located on the rear side of the device. For the meaning of
individual elements refer to legend Fig. 5, for operation of the pyrometer see section 7.

Fig. 5

Operating and display elements

Operating and display element

Meaning

1 Push button for pilot light

switch on/ off the pilot light

(optional through lens sighting)

2 Potentiometer

[emissivity]

adjustment of emissivity from 0.1 … 1

4.1.3

Connections and interfaces

The 5-pole plug to connect the instrument is at the rear side of the unit (see Fig. 66). The
contacts of the 5-pole plug are arranged as follows:

Plug pin

Meaning

1

Supply voltage +24 V (DC or AC)

2

Supply voltage 0 V (DC or AC)

3

ground output

4

Analogue output [0 (4) … 20] mA

5

empty