7 operation of the pyrometer, 1 setting the vario optic, Operation of the pyrometer – LumaSense Technologies IGA 740 User Manual

Page 22: Setting the vario optic, Fig. 10, Eration of the pyrometer

Advertising
background image



Operation of the pyrometer

18

KLEIBER

740

7

Op

7.1

Setting the Vario optic

Adjust the necessary measuring distance with the help of the vario optic as follows:

1. If necessary switch on the pilot light by pressing the push button at the rear side.

2. Loosen the ring nut (item 1, Fig. 10) by turning it in a counter-clockwise direction.

3. Move the vario optic (item 2, Fig. 10) forwards and/or backwards, in order to adjust

the pyrometer to the required measuring distance.
Note:

eration of the pyrometer

The size of the pilot light is identical to the size of the measuring field. The measuring
field may not be larger than the object to be measured at the adjusted distance!

4. After adjusting the measuring distance, lock the annular nut by turning it in a

clockwise direction.

The vario optic is thus adjusted to the required measuring distance.

Fig. 10

Setting the vario optic


a*

1

in mm

450

600

800

1000 1200 1400 1600 1800 2000 2500 3000

M*

2

in mm

2.5

4.0

6.0

8.0

10.5

11.5

13.0

14.0

15.0

20.0

28.0

*

1

Distance from optics front edge

*

2

Spot size diameter

Advertising