20

Power supply

The scale is powered by a mains unit or by a rechargeable batterie. The scale works in-
dependently of the mains, which is very handy for mobile use.
The LC display and the evaluation electronics have a low power consumption – 2000
weighing operations can be performed with a fully charged battery. Once battery voltage

is no longer adequate, the battery symbol

first appears in the display to indicate

that battery power is dropping. If the batterie is not charged, battery voltage will ultimately

not be adequate for error-free measuring. In this case,

bAtt

appears in the display and

weighing is no longer possible.
The rechargeable batterie is charged by

the mains unit supplied. For this purpose,

the mains unit needs to be connected to

the mains for at least 24 hours.
Please use only the mains unit sup-
plied or mains units from the seca

range of accessories. Before plugging

the unit into the socket, check wheth-
er the mains voltage marked on the

mains unit matches local mains volt-
age.

4. Setting up and aligning the scale

– Place the scale on a level surface.
– Fold the column upwards. Secure the

column by tightening up the knurled

screw.

– Undo the locknuts using the four foot

screws (1).

– Align the scale by undoing the foot

screws. The air bubble in the spirit level

must be right in the centre of the circle

(2).

– Then re-tighten the locknuts (3).
– Fold down the ramps required.
– Ensure that only the feet of the scale

are in contact with the floor. The scale

may not be in contact at any other

point.

Important!

The alignment of the base of the scale must be checked and corrected if

necessary every time the location of the scale is changed.

1.

2.

3.

Libelle

spirit level