AM1730 Secondary Reference Platinum Resistance Thermometer

User’s Guide

4/10

2 Specifications

2.1 Specifications

Temperature Range

1730-12: -200°C to 420°C

1730-9: -60°C to 300°C

Resistance at 0 °C

Nominal 100 Ω

Temperature Coefficient

0.003925 Ω/ Ω/°C

Accuracy

±0.025°C at -196 °C

±0.015°C at 0.01 °C

±0.035°C at 420 °C

Drift

±0.01°C at 0 °C after 500 hours at 420 °C

Short Term Stability

±0.007°C

Thermal Shock

±0.005°C after 10 times thermal cycles from minimum to

maximum temperatures

Hysteresis

<=0.005°C

Self-heating

50 mW/°C

Response Time

9 seconds for 63% response to step change in water moving

at 3 feet per second

Measurement Current

0.5 mA or 1 mA

Sensor Length

32 mm

Sensor Location

5 mm from tip

Insulation Resistance

>1000 MΩ at room temperature

Sheath Material

Inconel

tm

Dimension

1730-12: 0.25 inch X 12 inch (6.35 mm X 305 mm)

1730-9: 0.187 inch X 9 inch (4.75 mm X 229 mm)

External Leads

Teflon

tm

–insulated copper wire, 4 leads, 2.5 meters

Handle Dimension

1730-12: 15mm (OD) x 65mm (L)

1730-9: 10mm (OD) X 50 mm (L)

Handle Temperature Range

-50°C to 180°C

Calibration Options

1730-12-T, PRT with NIST traceable calibration and data

1730-9-T, PRT with NIST traceable calibration and data