AM1730 Secondary Reference Platinum Resistance Thermometer

User’s Guide

8/10

5 Troubleshooting

5.1 Troubleshooting

If the PRT functions abnormally, it could be caused by several

possible problem conditions that are described in this section. Try

the solutions recommended and if the problems are still not solved,

contact manufacture for warranty or repair service. Be sure to have

the model number and serial number of your PRT available.

5.2 Problem Causes and Solutions

a. R0/Rtpw becomes higher significantly. This is likely caused

by mechanical shocks. The PRT should be annealed to

release the stress of platinum wires and to recover R0/Rtpw

value. Measure R0/Rtpw of the PRT after annealing to verify.

b. R0/Rtpw unstable during the measurement. This is likely

caused either by bad connections or sensor coils short. Check

the connections first and if the connections are good, the PRT

may be damaged.