Additel 949 Hydraulic High Pressure Calibration Pump User Manual User Manual

Specification, Troubleshooting

Advertising
background image

Hydraulic Schematic

1-

Reservoir shut off valve

2-

Over-flow reservoir

3-

Positional Autoclave F-250-C,

9/16"-18UNF female

(

)

or customized female connector

4-

Reservoir cover vent valve

5-

Reservoir

6-

Liquid drain valve

7-

Pre-pressurization shut off valve

8-

High-pressure Pressurization and

fine pressure adjustment handle

9-

Pre-pressurization handle

Warnings and cautions

Views and Hydraulic Schematic

Hydraulic Ultra-High Pressure Test Pump User's Manual

[Version number:1306V01]

949

4

6

1

3

2

7

8

9

5

Specification

Operate the pump in the rated pressure range

,500 psi

,

DO NOT over the safety pressure

40,000 psi

.

37

Pressure range:

Temperature:
Adjusting fineness:
Safety pressure:
Pressure media:
Size:

Weight:

12.5 psi ( 0.85 bar ) vacuum to

37,500 psi ( 2500 bar ) positive pressure

5 ~ 50)℃

psi ( 1 mbar )

< 40,000 psi

Height: 10.43'' ( 265 mm )

Base: 20.71'' ( 526 mm ) x 9.65'' ( 245 mm )

35.5 lb ( 16 kg )

0.015

Diethylhexyl Sebacate

Close all valves and handles of the pump, and tighten the plugs
on the connectors during transportation.

Always keep the reservoir cover vent valve

#4

open during

operation.

Do not over tighten the valves, connectors and handles to avoid
any damage.

Change media immediately if it is contaminated.

Keep media level between 1/4 and 3/4 of the liquid reservoir filled.

Keep the threads clean and lubricious, and remove any dirt on
threads.
Used by Trained Personnel only.

Any security problems or damages caused by incorrect operations
are beyond Additel's responsibility.

Troubleshooting

Hard to pressurize

through turning

Pre-pressurization

handle (#9)

Problems

Causes

Solutions

Hard to pressurize through

turning high-pressure

pressurization handle (#8)

Hard to reach high vacuum

Hard to turn the

valves or handles

Step 1: Turn pre-pressurization handle (#9) clockwise fully, close pre-pressurization shut off valve (#7),

open reservoir shut off valve (#1).

Step 2: Turn pre-pressurization handle (#9) anticlockwise fully, close reservoir shut off valve (#1).
Step 3: Open pre-pressurization shut off valve (#7), pressurize the system.
Step 4: Repeat step 1 to 3.

Reservoir shut off valve (#1) is not closed

The media is not enough

The pre-pressurization shut off valve (#7) is not closed completely

The connector of DUT is not matched to connector

Pre-pressurization shut off valve (#7) is closed

Too much air in the pump

Reference gauge or devices under test (DUTs) are not connected tightly

Hard to pressurize

large-cavity DUT

Close shut off valve (#1)

Fill more media, and keep media level between 1/4 and 3/4 of the liquid reservoir filled

Close pre-pressurization shut off valve (#7)

Use proper adapter

Open pre-pressurization shut off valve (#7)

Turn the handle anticlockwise to vacuum the pump, open reservoir shut off valve (#1) to have air released

Check connector, replace the PTFE type, re-tight using a wrench

Close pre-pressurization shut off valve (#7), open reservoir shut off valve (#1), turn pre-pressurization handle
(#9) clockwise fully, close reservoir shut off valve (#1), and then open pre-pressurization shut off valve (#7)

Do not close shut off valves and handles too tightly

Lubricate the threads

It is normal. Please use high-pressure pressurization handle to increase/reduce the pressure.

Too tight in last operation

Lack of lubrication on threads

Hard to turn pre-pressurization handle (#9) at high pressure

Large-cavity of DUT, and pressurize the system too slow

Too much air in the pump

Always tighten the gauges using two wrenches .

1

7

9

8

5

#4 valve is not open

Open the #4 valve

Pressure gauges do not

reach to zero

Remark

.

If local atmosphere pressure is 1 bar, the vacuum can reach to 0 85 bar;

If local atmosphere pressure is P, the vacuum can reach to

bar.

( P 85% )

#4 valve is not open

Open the #4 valve

When the system pressure is greater than

,

the re-pressurization shut off valve(#7)

pressurize the system.

10,000 psi you must

close

p

and then

Advertising