Be f g, Basic operation connection, Maintenance – Additel 949 Hydraulic High Pressure Calibration Pump User Manual User Manual

Page 2: Remark, Pressurizing procedure depressurizing procedure, Fill media drain and clean, Degas, Pressure gauge, Pressure gauge bar

Advertising
background image

DIGITAL PRESSURE CALIBRA

TOR

Units

A

0

25

50

75

100

0

25

50

75

100

bar

bar

A

SW

V

mA

COM

24V

SW

V

mA

COM

24V

SW

V

mA

COM

24V

SW

V

mA

COM

24V

Basic Operation

Connection

dial gauge

pressure switch

pressure transmitter

flexible hose

D

B

E

F

G

A

B

Maintenance

Pressure Gauge

Pressure Gauge

Bar

32

64

96

128

160

C

②open

③close

⑤open

Pressure Gauge

Pressure Gauge

Bar

32

64

96

128

160

Pressure Gauge

Pressure Gauge

Bar

32

64

96

128

160

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

bar

bar

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

bar

bar

Pressure Gauge

Pressure Gauge

Bar

32

64

96

128

160

Pressure Gauge

Pressure Gauge

Bar

32

64

96

128

160

③close

①close

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

0

25

50

75

100

bar

bar

Pressure Gauge

Pressure Gauge

Bar

32

64

96

128

160

Pressure Gauge

Pressure Gauge

Bar

32

64

96

128

160

②open

④open

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

bar

bar

Pressure Gauge

Bar

0.8

0.6

0.4

0.2

0

-1

Pressure Gauge

Bar

0.8

0.6

0.4

0.2

0

-1

②close

④open

①open

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

bar

bar

Pressure Gauge

Bar

0.8

0.6

0.4

0.2

0

-1

Pressure Gauge

Bar

0.8

0.6

0.4

0.2

0

-1

①close

DIGITAL PRESSURE

CALIBRATOR

Units

A

0

25

50

75

100

bar

bar

Pressure Gauge

Bar

0.8

0.6

0.4

0.2

0

-1

Pressure Gauge

Bar

0.8

0.6

0.4

0.2

0

-1

②open

Pressurizing procedure

Depressurizing procedure

Always keep valve(#4)

open during operation

④a few turns

Pre-pressurizing

high-pressure

Pressurizing

Note: Skip step

for dial gauge

calibration in a narrow

pressure range

Depressurizing

Depressurizing

③Fully turn

⑤Fully turn

②Vacuum

①back to atmosphere

Open to atmosphere (Venting)

and Zeroing

Vacuum procedure

Back to atmosphere

②Fill media

①Open reservoir
cover vent valve (#4)

Open drain valve (#6)

Collect media using a container

Open reservoir cover

vent valve (#4)

Remove the
reservoir(#5)

Unscrew with

hex wrench

Fill media

Drain and Clean

Remark:

A Additel has made a concerted effort to provide
complete and current information for the proper
use of the equipment. The product specifications
and other information contained this manual are
subject to change without notice.
B

:

: Above pictures are just for reference.

Additel Corporation, USA

www.additel.com

Note:

Gauge positions are exchangeable. If only one device under test (DUT) is connected, the third connector on pump should be closed by a plug provided.

② Connect the gauges on pump by fingers, tighten the connectors using two wrenches.

Note: Please degas after filling media liquid,

in case hard to generate pressure.

Degas