Am-ldo luminescent dissolved oxygen sensor, Technical data dimensions ordering information, Related products – Aquametrix AM-LDO Luminesescent Dissolved Oxygen User Manual

Page 2: Dimensions

Advertising
background image

Measuring Range

4 ppb (0.5% air sat.) to 24 ppm (300%)

Wetted Materials

316 SS, EPDM, silicone

Accuracy at 25

0

C

1 ± 0.05%-vol; 21 ± 0.2%; 50 ± 0.5%

Storage Temperature

-10 to 50

0

C

Detection limit

0.1 % - vol

Operation Temperature

-10 to 60

0

C

90% Response Time <30 sec at 25

0

C

Pressure Range

-1 to 12 bar / 174 psi

Drift at Room Temp. < 0.2%-vol oxygen per week at 30

0

C

Wetted materials

CPVC

Power input

7-30 VDC. (0.6 W)

Not resistant to

Chlorine, Chloroform,

Toluene, Acetone

Analog Output

4-20 mA scalable. Tem-

perature compensated

Digital Output

Modbus RTU over RS485

Polarization voltage 5-10 min for self heating

Enclosure Type

IP 65

Transmission Rate

4800 to 115,000 bd

Connector

VP 8.0 Variopin con-

nector with PG 13.5

Cross sensitivity

None (inc. CO

2

, H

2

S, SO

2

)

Mounting

PG 13.5 - thread, IP68

AM-LDO Luminescent Dissolved Oxygen Sensor

Technical Data

Dimensions

Ordering Information

www.WaterAnalytics.net

978-749-9949

Toll free - 855-747-7623 (855-pHProbe)

Related Products

2200D Dissolved Oxygen Controller with P91 Clark style sensor
R655-X Series of 4-20 mA output ORP differential probes

AM-LDO Sensor w 4-20 mA output

AM-LDO-A

AM-LDO Sensor w Modbus RTU

AM-LDO-D

Replacement Cap

AM-LDO-Cap

USB-RS485 Modbus Converter

AM-LDO-CONV

Dimensions

Function

4-20 mA PS (7-30

VDC)

Ground NC

NC

RS485A 485B

Housing

VP Pin

A

B

C

D

E

F

G

H

Shield

Color

Yellow

Green

Red

Blue

Brown

White

Grey

Pink

Yellow/Green