COSA Xentaur Xentaur LPDT User Manual

Page 17

Advertising
background image

LPDT User’s Manual

page 10

Troubleshooting unexpected readings (table spans 2 pages)

For non-sensor related problems (e.g. no reading on instru-
ment) refer to section 3.5

Symptom

Possible Cause

Reading is not

changing

Condensation in sample system.

Slow Response

1. Water vapor in the system.
2. Flow rate too low.
3. Sample pipe too large and/or too long.
4. Unsuitable sample pipe material.
5. Leaks
6. Hygroscopic materials in sample system

Dry Reading

SpanCheck™ wrongly set, or faulty sensor.

Wet Reading

Leak in system or use of unsuitable pipe.

Comparison of readings with manual cooled-
mirror instrument.

Display Shows

6$7

Prolonged exposure to wet gas.

Display Shows

6+5

1. Instrument Failure

2.Short circuited sensor.

Display Shows

1. Instrument failure.

231

2. Open circuit on sensor

.

Advertising