Caution, Equitest 5071 – HT instruments EQUITEST5071 User Manual

Page 12

Advertising
background image

EQUITEST

5071

EN - 10

6.1.1. Calibration of measuring cables

1. Insert the black cable and the blue cable into the relevant input terminals B1 and B4 of

the instrument.

Fig. 3: Terminal connection during calibration procedure

2. If, for the measurement to be carried out, the length of the cables provided should be

insufficient, extend the blue cable.

3. Insert two alligator clips into the cable terminals.
4. Short-circuit the measuring cables’ ends, taking care that the conductive parts of the

alligator clips have good contact (see Fig. 3)

5. Press the F2 key. The instrument calibrates.

CAUTION

Never disconnect the terminals from the measuring points when the
instrument shows the message "MEASURING…".

LOW

 30.11.12


----


R+ R-
----

 ----

---mA ---mA


AUTO 0.11

This numerical
value indicates
that calibration has
been carried out

FUNC

CAL

6. At measurement end, the instrument gives out a double sound, to indicate that

calibration has been correctly carried out. The value of calibration is updated and
displayed above the F2 key. This value shall remain saved also after the instrument
has been switched off.


CAUTION

 The instrument calibrates the measuring cables only if their resistance is

lower than 5

 Before carrying out any measurement, always make sure that calibration

is referred to the cables currently used. In a continuity test, if the purified
resistance value (i.e. the value of resistance minus the calibration offset
value) is negative, the symbol “CAL” flashes on the display.


7. To delete the calibration parameter it is necessary to carry out a calibration

procedure with a lead resistance higher than 5

 (e.g. with open leads). When

carrying out a deletion, the symbol “- - -“ is shown on the display above the F2 key.