HT4014-HT4016

EN - 12

4.3.3. AC Current measurement

WARNING

• Make sure that all the test leads are disconnected from the meter's

terminals for current measurement.

• When measuring current, any strong current nears or closes to the

clamp jaws will affect the accuracy.

• The instrument is not available for non-sine wave AC signal.

Correct

Incorrect

Fig. 6: Use of clamp during AC current measurement

1. Select “

~A

” position.

2. Open the clamp and put the tested conductor in the center of the clamp jaw (see

paragraph 4.1.2.), considering the connection showed in Fig. 6.

3. The current value will be indicate on the display with automatic detection of the

appropriate range.

4. Pressing the R-H /

key (HT4016 only) you can select manually the measurement

ranges the measurement ranges in cyclic order. Press the R-H /

key more than 1

seconds at least to come back to AUTOMATIC range selection.

5. The "

O.L

" symbol means that the measured quantity is higher than the selected

range. Press the R-H /

key to select a higher range.

6. If the reading is difficult, press the D-H key to hold the obtained value. To exit from this

function press D-H key again.

7. Press MAX/MIN key to activate the saving of Maximum and/or Minimum values of

parameters. Press MAX/MIN key more than 1 second to exit function (see chapter
4.2.5. for details).

WARNING

As the clamp has internal filter it could happen that the meter takes few
seconds to reach 0 value on the display. This is not a defect, indeed the
user can proceed with a test as the temporary displayed value will not be
added to the measured value.