Warning, Correct incorrect – HT instruments HT4012 User Manual

Page 36

Advertising
background image

HT4012

EN - 11

4.3.3. AC Current measurement

WARNING

• Make sure that all the test leads are disconnected from the meter's

terminals for current measurement.

• When measuring current, any strong current nears or closes to the

clamp jaws will affect the accuracy.

• The instrument is not available for non-sine wave AC signal.

Correct

Incorrect

Fig. 6: Use of clamp during AC current measurement

1. Select “

~A

” position.

2. Open the clamp and put the tested conductor in the center of the clamp jaw (see

paragraph 4.1.2), considering the connection showed in Fig. 6.

3. The current value will be indicate on the display with automatic detection of the

appropriate range.

4. Pressing the R-H key you can select manually the measurement ranges the

measurement ranges in cyclic order. Press the R-H key 2 seconds at least to come
back to AUTOMATIC range selection.

5. The "

O.L

" symbol means that the measured quantity is higher than the selected

range. Press the R-H key to select a higher range.

6. If the reading is difficult, press the D-H key to hold the obtained value. To exit from this

function press D-H key again.

wa

WARNING

As the clamp has internal filter it could happen that the meter takes few
seconds to reach 0 value on the display. This is not a defect, indeed the
user can proceed with a test as the temporary displayed value will not be
added to the measured value.