HT6-HT8-HT9

EN - 7

5. INSTRUMENT FUNCTIONS

5.1. VOLTAGE

MEASUREMENT

AND POLARITY INDICATION

WARNING

 The maximum input AC and DC voltage is 690V. Do not measure

voltages exceeding the limits given in this manual. Exceeding these
limits could result in electrical shocks to the user and damage to the
instrument.

 The Dangerous voltage led (see Fig. 2 – no. 5) lights up and the buzzer

sounds, in case of voltage between terminals >50V AC or >50V DC,
even in case of low battery charge or when no batteries are installed
(without batteries the buzzer does not sound).

 After a lamp test it is necessary to wait 5 seconds before being able to

perform a voltage measurement.

Fig. : 3: Voltage measurement

1. Connect the measuring leads L1 and L2 to

the object to be tested. Measurement is only
carried out by contact between the leads
and the metal parts of the conductors

.

2. The instrument automatically switches on and

the measured voltage value is indicated when
the leds turn on and, for models HT8 and
HT9, also on the LCD display (see Fig. 2 –
nos. 1 and 8).

3. While measuring, the instrument could also

show leds

or

on. This is no indication

of malfunction.

Polarity is indicated as follows

(see

Fig. 2 – no.

7)

:

 AC: leds + and – 12V are on;

 +DC: led +12V is on (lead L2 connected to

positive pole);

 -DC: led -12V is on (lead L2 connected to

negative pole).


In case of measurements in environments with
poor light, press the Torch button (see Fig. : 1 –
no. 7) to activate the white LED pointer.