Ht603, User manual – HT instruments HT603 User Manual

Page 2

Advertising
background image

ENG

HT603

USER MANUAL

1. PRECAUTIONS AND SAFETY MEASUREMENTS

This instrument complies with safety Standard IEC/EN61010-1
related to electronic measuring instruments. For your own safety and
to avoid damaging the instrument follow the procedures described in
this instruction manual and read carefully all notes preceded by this

symbol

CAUTION

• Avoid measuring in humid or wet places

• Avoid measuring in rooms where explosive

gas, combustible gas, steam or excessive dust
is present

• Keep you insulated from the object under test

• Do not touch exposed metal parts such as test

lead ends, sockets, fixing objects, circuits etc.

• Avoid doing that if you notice anomalous

conditions such as breakages, deformations,
fractures, leakages of battery fluid, blind display
etc.

• Be careful when measuring voltages exceeding

20V to avoid risks of electrical shock


The following symbols are used in user manual and on the meter:

CAUTION - refer to the instruction manual - an
improper use may damage the instrument or its
components

Double insulated meter

Danger high voltage: risk of electric shock

AC voltage

DC voltage

CAUTION: this symbol indicates that equipment, its
accessories and battery shall be subject to a
separate collection and correct disposal

1.1 Preliminary instructions

CAUTION

• This instrument has been designed for use in

environments with pollution degree 2

• It can be used for VOLTAGE measurements

on installations with overvoltage category CAT
III 300V and CAT II 600V

• Only the leads supplied with the instrument

guarantee compliance with the safety
standards in force. They must be under good
conditions and, if necessary, replaced with
identical ones

• Do not test or connect to any circuit exceeding

the specified overload protection

Do not effect measurements under
environmental conditions exceeding the limits
indicated in § 8

• Make sure that batteries are properly installed

• Before connecting the test probes to the

installation, make sure that the function
selector is positioned on the required
measurement

• Make sure that LCD and function selector

indicate the same function

• When changing the range, first disconnect the

test leads from the circuit under test in order to
avoid any accident

When the instrument is connected to
measuring circuits never touch any unused
terminal

• If during measurement the displayed values

remain constant check whether the HOLD
function is active

1.2 Measuring overvoltage category definitions
According to the IEC/EN61010-1 guidelines the circuits are divided
into the following measurement categories:
CAT IV is for measurements performed at the source of the low-

voltage installation

CAT III is for measurements performed in the building installation

CAT II is for measurements performed on circuits directly

connected to the low voltage installation

CAT I is for measurements performed on circuits not directly

connected to MAINS

2. GENERAL DESCRIPTION

HT603 performs the following measurements:

• DC and AC voltage

• Resistance and continuity test

• Frequency and Duty Cycle

• Capacitance

• Diode test

All selectable by means of a 7-position function selector. Function
keys HOLD, SELECT and RANGE/REL

∆ are also available (see

§ 5.1, § 5.2 and § 5.3). The selected quantity is displayed with
indication of measuring unit and active functions. The instrument
is provided with an Auto Power Off function consisting in an
automatic switching off 30 minutes after last pressure on keys or
rotation of selector. To resume normal operation, switch on the
instrument again

3. PREPARATION FOR USE

3.1 Initial checks
This instrument was checked both mechanically and electrically
prior to shipment. All possible cares and precautions were taken
to let you receive the instrument under perfect conditions.
Notwithstanding we suggest you to check it rapidly (any damage
may have occurred during transport – if so please contact the
local distributor from whom you bought the item). Make sure that
all standard accessories mentioned in § 8 are included. Should
you have to return back the instrument for any reason please
follow the instructions mentioned in § 9

3.2 Supply voltage
The instrument is powered by 2x1.5V button batteries type LR44
included on meter. When batteries are low the symbol "

" is

displayed. To replace/insert batteries follow the instructions
indicated in § 7

3.3 Calibration
The instrument complies with the technical specifications
contained in this manual and such compliance is guaranteed for
12 months

3.4 Storage
After a period of storage under extreme environmental conditions
exceeding the limits mentioned in § 8 let the instrument resume
normal operating conditions before using it

4. INSTRUMENT AND DISPLAY DESCRIPTION


© Copyright HT ITALIA 2011 Release EN 1.01 - 19/12/2011

5. FUNCTION KEYS DESCRIPTION

5.1 HOLD key
By pressing HOLD key the measured value is hold on the display
where the symbol "HOLD" appear. Pressing again HOLD to disable
this function and resume normal operation

5.2 SELECT key

Press SELECT key to select the sub-functions relative to the Hz%,

and Hz% posiction of rotary switch


5.3 RANGE/REL

key

By pushing RANGE/REL

∆ key the manual mode is activated

(functions

, and

Ω) and the “AUTO” symbol disappears. Press

RANGE/REL

∆ key (only function

) to enable relative

measurement. The “REL

∆” symbol is shown and the value is zeroed

at display. Press again the key to exit from the function

5.4 Disable Auto Power OFF
For the disable of Auto Power OFF feature:
1. Switch off the instrument
2. Switch on the instrument by press and hold RANGE key

6. HOW TO PERFORM THE MEASUREMENTS

6.1 AC Voltage, Frequency and Duty Cycle measurements

CAUTION

Maximum input for AC voltage measurements is
600V. Do not take any voltage measurement
exceeding this limit in order not to risk electrical shock
or damaging the meter

1. Rotate the selector on

Hz% position. The “AC” and “AUTO”

symbols are shown at display. Press SELECT key for AC voltage
measurement selection (symbol “V” at display) and press
RANGE key for the selection of manual range (see § 5.3)

2. Connect the red and black test leads on the circuit under test.

The AC voltage value is displayed

3. If the message "OL." is displayed the maximum voltage value

was exceeded

4. Press SELECT key for the visualization of frequency value. The

symbol “Hz” is shown at display

5. Press

SELECT key for the visualization of Duty Cycle value. The

symbol “%” is shown at display

6. For Data HOLD feature see § 5.1


6.2 DC

Voltage

measurement

CAUTION

Maximum input for DC voltage measurements is
600V. Do not take any voltage measurement
exceeding this limit in order not to risk electrical shock
or damaging the meter

1. Rotate the selector on or m position. The “DC” and “AUTO”

(only ) symbols are shown at display. Press RANGE key for the

selection of manual range (only ) (see § 5.3)

2. Connect the red test lead (positive pole) and black test lead

(negative pole) on the circuit under test. The DC voltage value is
displayed

3. The symbol "-" on the instrument display indicates that the

voltage has opposite direction compared to the previous
connection

4. If the message "OL." is displayed the maximum voltage value

was exceeded

5. For Data HOLD feature see § 5.1

6.3 Resistance, Continuity test and Diode test measurements

CAUTION

Before taking resistance measurements on circuit
remove power from the circuit under test and
discharge all capacitors


1. Rotate the selector onto

position. The “M

Ω” and “AUTO”

symbols are displayed. Press RANGE key for the selection of
manual range (see § 5.3)

2. Connect the test leads to the circuit under test. The resistance

value is displayed

3. Press SELECT key for Continuity Test selection (symbol “ ” at

display). The buzzer sounds when resistance values are lower
than 50

4. Press SELECT key for DiodeTest selection (symbol “ ” at

display). Connect the red lead to the positive side (anode) of the
diode and the black lead to the negative side (cathode). The
meter displays the diode voltage typically within 0.4 ~ 0.9V.
Reverse the connections and measure the voltage across the
diode again. The message “OL." on the display corresponds to a
correct junction

6.4 Frequency and Duty Cycle measurements

CAUTION

Do not perform measurements on circuits with a
voltage exceeding 600V AC/DC to avoid possible
shock hazard


1. Rotate the selector onto Hz% position. Press SELECT key for

Frequency or Duty Cycle measurement selection. The “Hz” or “%”
symbols are shown at display. Connect the test leads to the circuit
under test. The frequency value (in kHz or MHz) or the percentage
value are shown at display


6.5 Capacitance measurement

CAUTION

Before taking any in circuit or capacitance
measurement, remove power from the circuit being
tested and discharge all capacitors

1. Rotate the selector onto ”

” position. Press RANGE/ REL

∆ key

for “nF” capacitance measurement. The value is zeroed at display

2. Connect the test leads to the capacitor terminals taking care of the

positive polarity (red test lead) and negative polarity (black test
lead). The capacitance value will be displayed

3. If the message "OL." is displayed the maximum capacitance value

was exceeded

7. BATTERY REPLACEMENT

When the "

" low battery indication is displayed the batteries must

be replaced as follows:

1. Switch

OFF the meter

2. Remove the fixing screw and the battery compartment cover
3. Remove batteries and replace with new ones of the same type
4. Replace

battery

compartment cover

8. TECHNICAL SPECIFICATIONS

The accuracy is referred to 23°C± 5°C, <80%RH
AC Voltage

Input impedance: 10M

Ω, <100pF; Overload protection: 600V rms

DC Voltage

Input impedance: 10M

Ω, <100pF; Overload protection: 600V rms

Resistance and Continuity test

Frequency

Sensitivity: 1.0Vp-p (square wave) ; Vmax: 5.0Vp-p, 240V (50/60Hz)

Duty Cycle

Sensitivity: 1.0Vp-p 30%

≤Duty≤70% (square wave) ; Vmax: 5.0Vp-p

Capacitance


Meas. Range

Resolution

Frequency

Accuracy

400.0mV 0.1mV

50 ÷ 500Hz

±(1.5%rdg+5dgt)

4.000V 0.001V

±(0.9%rdg+5dgt)

40.00V 0.01V

400.0V 0.1V

600V 1V

Meas. Range

Resolution

Accuracy

400.0mV 0.1mV

±(0.7%rdg+5dgt)

4.000V 0.001V

±(0.6%rdg+2dgt)

40.00V 0.01V

400.0V 0.1V

600V 1V

±(0.7%rdg+5dgt)

Meas. Range

Resolution

Buzzer

Accuracy

400.0

Ω 0.1Ω

<50

(range
400

Ω)

±(0.9%rdg+5dgt)

4.000k

Ω 0.001kΩ

±(0.9%rdg+2dgt)

40.00k

Ω 0.01kΩ

400.0k

Ω 0.1kΩ

4.000M

Ω 0.001MΩ

±(1.5%rdg+5dgt)

40.00M

Ω 0.01MΩ

Meas. Range

Resolution

Accuracy

0.001Hz ÷5MHz

0.001Hz

÷ 0.001MHz

±(0.3%rdg+5dgt)

Meas. Range

Resolution

Accuracy

0.1% ÷ 99.9%

0.1%

±(0.5%rdg+3dgt)

Meas. Range

Resolution

Accuracy

50.00nF 0.01nF

±(5.0%rdg+0.2nF)

500.0nF

0.1nF

±(2.9%rdg+5dgt)

5.000µF

0.001µF

50.00µF

0.01µF

100.0µF

0.1µF

Safety: IEC/EN61010-1
Measurement category: CAT III 300V , CAT II 600V, Pollution degree: 2
Insulation: double insulation
Compliance with 2004/108/CE (EMC) and 2006/95/CE (LVD) Directives
Display: LCD, 4dgt, 5000 points
Sampling rate: 3 times/s
Diode test: test current 1.1mA, open voltage: 1.5VDC (max)
Over range indication: “OL.” or “-OL.” symbols at display
Power supply: 2x1.5V button batteries type LR44
Auto Power OFF: after 30 minutes of idleness
Reference temperature/humidity: 23°C± 5°C (73°F ± 41°F) ; <80%RH
Working temp.: 0

÷30°C <80%RH ; 30 ÷40°C <75%RH ; 40 ÷50°C <45%RH

Storage temperature/humidity: -20

÷ 60°C (-4°F ÷ 140°F) ; <80%RH

Max height of use: 2000m (6562ft)
Dimensions (LxWxH): 112x56x12mm (4x2x0.5in)
Weight (with batteries): 115g (0.2lv)
Included accessories: test leads, batteries (fitted), user manual

9. SERVICE

9.1 Warranty conditions
This instrument is guaranteed for one year against material or
production defects, in accordance with our general sales conditions.
During the warranty period the manufacturer reserves the right to
decide either to repair or replace the product. Should you need for
any reason to return back the instrument for repair or replacement
take prior agreements with the local distributor from whom you
bought it. Do not forget to enclose a report describing the reasons for
returning (detected fault). Use only original packaging. Any damage
occurred in transit due to not original packaging will be charged
anyhow to the customer

9.2 Service
Should the instrument not work properly, make sure that batteries are
correctly installed and working before contacting your distributor, also
check the test leads and replace them if necessary. Should you need
for any reason to return back the instrument for repair or replacement
take prior agreements with the local distributor from whom you
bought it. Do not forget to enclose a report describing the reasons for
returning (detected fault). Use only original packaging. Any damage
occurred in transit due to not original packaging will be charged
anyhow to the customer. The manufacturer will not be responsible for
any damage to persons or things