User manual, Ht70 voltage and phase sequence detector, Safety precautions and procedures – HT instruments HT70 User Manual

Page 3: Caution, Introduction, Features, Description, Phase sequence indication and conformity test, Voltage detection test, Batteries replacement

Advertising
background image

USER MANUAL

HT70 voltage and phase sequence detector

1. Safety precautions and procedures

Please follow the herewith indications before and during the performance
of any test with the instrument:

CAUTION

 An improper use can damage the instrument and/or

its components or injure the operator

 Avoid using the instrument in rooms where explosive

gas, combustible gas, steam or excessive dust are
present

 Do not perform any test if you notice anomalous

conditions such as breakages, deformations,
fractures, leakages of battery liquid on meter

 Do not perform any test exceeding the declared

voltage conditions

 Do not perform measurements under environmental

conditions exceeding the declared limits

 Do not perform any voltage detection with operator’s

hand beyond the hand safety protection

2. Introduction

HT70 meter can perform AC voltage detection and phase sequence
rotation with visible LED and acoustic buzzer indications

3. Features

 AC voltage detection on conductive and insulated parts
 Phase sequence indication and conformity phase test
 Test on outlets, cables and/or electrical appliances
 Visible LED and acoustic buzzer indications
 Practical pen clip and very small size
 Auto

Power

OFF

 Meter suitable for any electrical installer

4. Description


5. Phase sequence indication and conformity test

CAUTION

 The test results are not granted on systems with

voltage values out of the 100V

 1000V AC range and

50

 60Hz frequency range.

 To perform tests correctly check that L1, L2 and L3

phase terminals are not so close to each other in
order to avoid any influence of their signals

Do not perform tests with the operator’s hands

beyond the hand safety protection and keep not
insulated wires beyond the CAT IV reference mark
(see part 2 of picture)

1

Voltage detection sensor

2

Reference mark for CAT IV measurements

3

Hand safety protection

4 ON/OFF

key

5 Pen

clip

6

Internal section Green / Red LED

7

Fit part for battery cover sleigh

8

Battery cover sleigh

1.

Press ON/OFF key for at least 1s. The green and red LEDs turn on in
a sequence and the buzzer is active for a while. After that, the meter
gets ready to start entering a standby mode and the green LED is
blinking continuously

2.

Take the meter and position the sensor on L1 wire of a three phase
plant (on metal or insulated part) for at least 5s in order to activate
the phase sequence test. The red LED and the buzzer turn on during
detection of correct voltage and frequency. The red LED along with
the buzzer continue blinking until the correct synchronization is
performed on L1 phase

3. Disconnect the sensor from L1 phase and wait for the turning off of

both red LED and buzzer before positioning it on L2 phase

CAUTION

If you wait for more than 10s between the first and the
second measurement, the green and red LEDs as well
as the buzzer blink in a sequence for some seconds,
then the meter goes back to its standby mode and the
test shall be repeated

4. Position and hold the sensor on L2 wire of a three phase plant (on

metal or insulated part) for at least 5s. The red LED and the buzzer
turn on in order to detect correct voltage and frequency

5.

At the end of the signal detection on L2 phase the below results are
given by the meter:
 Green LED turns on and buzzer intermittently sounds for some

seconds before meter goes back to its standby mode

Correct

phase sequence indication

 Intermittent working of red LED and buzzer for some seconds

before meter restores its standby mode

Incorrect phase

sequence indication

5.1 Phase conformity test

1. Repeat points 2) 3) and 4) on the L1 phases of both three phase

systems in case of conformity test. At the end of test the below
results are given by the meter:
 Green LED blinks and buzzer sounds intermittently for some

seconds before meter restores its standby mode

Correct

phase conformity

 Red LED blinks and buzzer sounds intermittently for some

seconds before meter goes back to its standby mode

Incorrect phase conformity

6. Voltage detection test

1. Turn on the meter by pressing ON/OFF key for at least 1s. Waiting

for the standby mode when the green LED is blinking

2.

Position the sensor close to or in contact with the point under test for
a time <5s, then verify that the red LED is on and buzzer sounds
continuously as evidence of voltage detection. Do not perform test
with the operator’s hand beyond the hand safety protection and
position not insulated wires beyond the CAT IV reference mark
(see part 2 of picture

3.

Press ON/OFF key for at least 3s to turn off the meter

7. Batteries replacement

Follow the herewith steps to replace internal batteries:

1.

Remove the meter from any voltage source

2. Perform a small pressure on battery cover sleigh (see part 8 of

picture)

3.

Shift the battery cover sleigh from the meter

4. Replace the batteries with new ones of the same type and pay

attention to keep to correct polarity indication

5.

Fit the battery cover sleigh to the meter

6.

Used batteries must not be disposed of as urban waste

CAUTION: this symbol indicates that the equipment and its
accessories shall be subject to a separate collection and
correct disposal.


8. Technical specifications

Voltage reference range: 100

 1000V AC to ground

Frequency reference range: 50/60Hz

Meter indications: visible LED and acoustic buzzer

 Safety:

IEC/EN61010-1

Category of measurement: CAT IV 1000V

Pollution degree: 2

Batteries: 2 x 1.5V alkaline type AAA IEC LR03

Autonomy: >9000 tests

Auto Power OFF: after 5 minutes of idleness

Working temperature: -10°C

 50°C (14°F  122°F)

Working humidity: -10

 40°C <75%HR ; 40  50°C <45%HR

14°F

 104°F <75%HR ; 104°F122°F <45%HR

Dimensions: 160 (L) x 26 (W) x 20(H) mm ; 6 (L) x 1 (W) x 1 (H) in

Weight (included batteries): 48g (0.1lb)