4.4. MEASUREMENTS - DESCRIPTION
4.4.1. DC voltage measurement

CAUTION

The maximum input for DC voltage is 600V. Don’t try to measure higher
voltages to avoid risks of electrical shocks or serious damages to the
instrument.

Fig. 3: DC voltage

measurement

1. Turn the selector on

.

2. Insert the black test lead in the COM jack and the red

probe in the V+/Ω jack (Fig. 3).

3. Connect the red probe and the black test lead

respectively to the positive and negative poles of the
circuit under test; the voltage value will be displayed
(automatic range detection).

4. Should you get the message "

O.L

" it means that the

detected voltage exceeds the limits which the
instrument can measure. In this case stop the
measurement by disconnecting the test leads from the
circuit under test to avoid risks for yourself and for the
instrument.

5. Should you get a symbol "-" it means that the detected

voltage has an opposite direction compared to the
connection shown in Fig. 3.

6. If you want to activate the HOLD function refer to

chapter 4.2.